Remote Operation 13-5 Trigger Commands
PowerXpert UG PN: 10585-00020 Rev. C 13-27
Trigger Delay
Trigger delay is the time in milliseconds between the trigger event and when the sensor starts taking readings.
The trigger delay can be either positive or negative. The trigger delay can be set from –5 ms to 10,000 ms with
a resolution of 0.01 ms.
If the delay is negative, the sensor starts taking the readings before the trigger occurs and the total capture
time includes the negative delay time. For example, for a capture time of 20 ms and a delay of –1 ms, the
length of the capture would be from –1 ms to 19 ms, given the trigger occurs at time, t = 0. The negative delay
can not be greater than or equal to the capture time. If the capture time conflicts with the trigger delay, the
trigger delay command returns an error.
If the delay is positive, the sensor waits for the set delay time after a trigger before it starts taking readings.
The sensor is unresponsive during the wait period and can not be aborted. The capture time is unaffected by a
positive trigger delay.
TRGDLY
Description: Sets the trigger delay in msec
Syntax: TRGDLY <trigger_delay> +LF
Return Value: OK or ERR
Remarks: <trigger_delay> is the trigger delay value in msec and must range from –5 ms to
10,000 ms with a resolution of 0.01 ms.
Compatible Sensor: MA24108A, MA24118A, MA24126A
TRGDLY?
Description: Gets the trigger delay in msec
Syntax: TRGDLY? +LF
Return Value: Trigger delay in ms
Compatible Sensor: MA24108A, MA24118A, MA24126A