Anritsu MA24106A Guia do Utilizador Página 169

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 224
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 168
Operational Testing for the MA24108A, MA24118A, and MA24126A 12-6 Linearity Test
PowerXpert UG PN: 10585-00020 Rev. C 12-17
Table 12-16.Measurement Results (24 GHz)
Row
#
Synthesizer
Power Level
Setting
(dBm)
Attenuation
in Reference
Arm
(dB)
A
Reference
Power
Measurement
(dBm)
B
= (A6–A5)
Correction
(dB)
C
= (A+B)
Corrected
Reference Power
Measurement
(dB)
Attenuation
in Test Arm
(dB)
D
MA241xxA
Measurement
(dBm)
E
= (C–D)
Difference
Calculation
(dB)
1+20 10 0
2+15 10 0
3+10 10 0
4+5 10 0
50 10 0
6 adjust per
Step 7e
00 10
7+5 0 0 10
80 0 0 10
9–5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):
Table 12-17.Measurement Results (26 GHz)
Row
#
Synthesizer
Power Level
Setting
(dBm)
Attenuation
in Reference
Arm
(dB)
A
Reference
Power
Measurement
(dBm)
B
= (A6–A5)
Correction
(dB)
C
= (A+B)
Corrected
Reference Power
Measurement
(dB)
Attenuation
in Test Arm
(dB)
D
MA241xxA
Measurement
(dBm)
E
= (C–D)
Difference
Calculation
(dB)
1+20 10 0
2+15 10 0
3+10 10 0
4+5 10 0
50 10 0
6 adjust per
Step 7e
00 10
7+5 0 0 10
80 0 0 10
9–5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):
Vista de página 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 223 224

Comentários a estes Manuais

Sem comentários