12-5 Frequency Response Test Operational Testing for the MA24108A, MA24118A, and MA24126A
12-4 PN: 10585-00020 Rev. C PowerXpert UG
12-5 Frequency Response Test
In this test the frequency response of the sensor is tested at one low power level against a reference sensor of
known measurement uncertainty. The reference sensor should be calibrated by a reputable standards
laboratory using instruments with low published measurement uncertainty values. To perform the
comparison, both sensors are used to measure the output power of a synthesizer with a high quality
attenuator, such as the 41KC-6, on the output. The attenuator improves the source match of the synthesizer by
lowering the mismatch ripples, thereby lowering the uncertainty in the comparison.
Test Procedure
1. Set up the equipment as follows (refer to Figure 12-1 for an illustration):
a. Connect the reference power sensor to the reference power meter using the appropriate cables.
b. Connect the USB cable between the personal computer with the PowerXpert application installed
and the MA24108A, MA24118A, or MA24126A power sensor under test.
c. Launch the PowerXpert application.
d. Turn the power on to all of the instruments and allow them to warm up for the amount of time
specified in their respective manuals.
e. Reset or Preset all of the instruments.
f. Configure the reference meter and sensor to measure a CW signal.
g. Perform a sensor Zero and a 1 mW reference calibration on the reference sensor and meter per the
manufacturer’s instructions.
h. Perform a low level Zero of the MA24108A, MA24118A, or MA24126A power sensor by
disconnecting it from the synthesizer, clicking the Zero button on the PowerXpert application, and
waiting for the Zeroing message to close.
i. Connect the attenuator to the output of the synthesizer with the appropriate adapter (if required)
to the output of the attenuator.
j. Set the synthesizer to +6 dBm and 50 MHz.
Index Description
1 Synthesizer
2 Reference Power Meter
3 Reference Power Sensor
4 K to N Adapter (if required)
5 Attenuator
Figure 12-1. Cal Factor Test Set Up (1 of 2)
MA24108A
MA24118A
MA24126A
1
2
3
4
5
6
7
3