DD
DR Set GPIB address
Syntax:
DR<val>
val:
1 to 30
Remarks:
Changes the device address when operating in HP emulation mode. The
power meter default address is 13.
EMUL GPIB emulation mode
Syntax:
EMUL <mode>
mode:
ML24XX (Anritsu ML2430A Series native mode)
HP436A (Hewlett-Packard)
HP437B (Hewlett-Packard)
HP438A (Hewlett-Packard)
ML4803 (Anritsu ML4803A Series)
Remarks:
Set the GPIB emulation to emulate other types of power meters. This com-
mand is available in any emulation mode, and resets the whole GPIB inter-
face when the emulation mode is changed.
When selecting GPIB emulation modes, the instrument configures itself to
the preset conditions of the instrument to be emulated. For example, when
selecting HP 438A emulation, the front panel menus pass through the pre
-
sets for the HP 437B (which presets sensor A to dBm) then selects HP
438A emulation (which presets sensor A to Watts).
FA auto average
Syntax:
FA
Remarks:
Automatic Filter on. Allows the system to automatically select the filter
used to reduce the jitter in the display.
Related
Commands:
FM, FH
FH average hold
Syntax:
FH
6-120 ML2430A OM