Página 1 - POWER METER

ML2430A SERIESPOWER METEROPERATION MANUALP/N: 10585-00001REVISION: KPRINTED: MARCH. 2000COPYRIGHT 2000 ANRITSUANRITSU LTD (EMD)RUTHERFORD CLOSESTEVENA

Página 2 - TRADEMARK ACKNOWLEDGMENTS

1-4IDENTIFICATIONNUMBERThe ML2430A Series ID number is affixed to the rear panel (see Figure3-2). Please use the complete ID number when ordering part

Página 3 - Safety Symbols

configuration combines sensors, the combination is done inlinear units. If the result of the combination produces anegative linear value and the displ

Página 4 - For Safety

*ESE? Return Event status register enable mask GPIB 488.2Syntax:*ESE?Remarks:Returned format: <unsigned character>When converted to an 8-bit bin

Página 5

Remarks:Places a single ASCII character '1' on the GPIB output queue when theconditions for the *OPC command are met. An operation is comple

Página 6 - Table of Contents

RelatedCommands:*RCL*SRE Setup service request enable registerGPIB 488.2Syntax:*SRE <val>val:8-bit maskRemarks:Sets the Service request enable r

Página 7

*TST? Self Test GPIB 488.2Syntax:*TST?Remarks:Performs a self test and returns 'PASSED' or 'FAILED.'NOTE: This command will restar

Página 8 - ML2430A OM iii/iv

same command as the one which changes to AUTO averaging, it will alsoupdate the USER averaging number.Example:AVG A, AUTO, 64This command will set the

Página 9 - General Information

AVGLL Auto low level averaging SENSORSyntax:AVGLL <s>,<mode>s:A or Bmode:OFFLOWMEDIUMHIGHRemarks:Sets the low level averaging window for t

Página 10 - INFORMATION

Syntax:BAUTS <state>state:ENABLE or DISABLERemarks:Enable/disable the battery auto power shut off.NOTE: Although GPIB is not available under bat

Página 11 - GENERAL SENSORS

NOTEIf the buffering enabled is set to OFF and '*OPC?' is used,the '*OPC?' will clear the output buffer of any previous re-sponse

Página 12 - SENSORS GENERAL

val:.07 to 150%+31.55 to -1.76dBRemarks:If the Cal factor source is set to manual, this is the calibration factornumber used.Example:CFCAL A, %, 99Set

Página 13 - Installation

1-6SENSORSThe following sensors, sensor options, and sensor accessories are avail-able for use with the ML2430A Series Power Meters:Power Sensors (–70

Página 14 - 2-2 ML2430A OM

Manual uses the CFCAL number itself.VGHz takes the frequency from the V/GHz input and uses it to look up thecalibration factor from the EEPROM in the

Página 15 - ML2430A OM 2-3

Remarks:Clears the cal factor table to one entry for 50 MHz at 100%, but does notclear the identity of the table. The cleared table is automatically s

Página 16 - RACK MOUNTING INSTALLATION

binary data:Same data as that recieved by CFURDRemarks:Loads binary data into the cal factor table. This command will automaticallysave the data to th

Página 17 - INSTALLATION RACKMOUNTING

This message can be manipulated to program a different table using theCFULD command.CFUSAV Cal factor table saveSENSORSyntax:CFUSAVRemarks:This comman

Página 18

Remarks:Returns a number indicating the cal factor table, or combination of tables,being used by the selected sensor. Possible returned values are:0 =

Página 19

CHCFG Channel input configuration CHANNELSyntax:CHCFG <c>,<config>c:1 or 2config:OFF,A, B, VA–B, B–AA/B, B/ARemarks:A, B, V = Sensor A, Se

Página 20

units:W (Watts)DBM (dB)DBUV (dBmV)DBMV (dBmV)Remarks:DBM 0dB is equal to 1mW readout modeW = Watts readout modeV = Volts readout mode. This selection

Página 21 - FASTENERS

RelatedCommands:GRDDT, SRCSPFRQ, SRCSTFRQ, SRCSTPWR, SRCSPPWRCURLK Link cursors in all graphic modesPROFILE SETUPSyntax:CURLK <state>state:ONOFF

Página 22 - 2-10 ML2430A OM

RelatedCommands:CVSPF, CVSTF, CVSTVQuery:CVSPV? <s>ReturnedString:CVSPV <s>,<val>CVSTF V/GHz calibration factor start frequencySENSO

Página 23 - ML2430A OM 2-11

Remarks:Sets the mode of the LCD backlight when under Battery power.ON = back light is ON all the timeOFF = back light is OFF all the timeTIMED = back

Página 24 - 2-12/2-12 ML2430A

SensorAccessoriesML2400A-20 1.5m SensorCableML2400A-21 0.3m SensorCableML2400A-22 3m Sensor CableML2400A-23 5m Sensor CableML2400A-24 10m Sensor Cable

Página 25 - Connections

ReturnedString:DCONT <val>DCONTD Set display contrast down by oneDISPLAYSyntax:DCONTDRemarks:Make the display lighter by lowering the contrast b

Página 26 - 3-2 ML2430A OM

B = Sensor B onlyA&B = Sensors A and B displayed at the same timeOFF = Turns the peak meter display off.The peak meter display range covers 12 dB.

Página 27 - ML2430A OM 3-3/3-4

RelatedCommands:DUTYSQuery:DUTY? <s>ReturnedString:DUTY <s>,<duty cycle>DUTYS Duty cycle stateSENSORSyntax:DUTYS <s>,<state

Página 28

ENTERR Entry Error beep SYSTEMSyntax:ENTERR <state>state:ON or OFFRemarks:Turns the user entry error warning beep On or Off.Query:ENTERR?Returne

Página 29 - Front Panel Operation

NOTESThe GPIB command error and GPIB execution error arealways enclosed within exclamation marks (!). If no errorshave been produced since the last ER

Página 30 - 4-2 ML2430A OM

e. Sensor OFFSETS are applied.f. Relative is applied if it is set to on before switching to FAST mode, and ifdisplay channel 1 is configured for a sin

Página 31 - OPERATION POWER-ON PROCEDURE

FROFF Frequency/Offset Display SYSTEMSyntax:FROFF <state>state:ON or OFFRemarks:This command turns on the top line information text displaying t

Página 32 - POWER-ON PROCEDURE OPERATION

GPRST Reset min/max graph PROFILE SETUPSyntax:GPRSTRemarks:When profile DATA HOLD mode is set to Min/Max points, this command isused to reset the min/

Página 33 - OPERATION SENSORMENU

ReturnedString:GRCP <state>GRDATA Display Graph Cursor DataPROFILE SETUPSyntax:GRDATA <state>state:ON or OFFRemarks:Display the graph curs

Página 34 - SENSOR MENU OPERATION

GRDRQ <channel_number>,<Cursor_1_dB>,<Cur-sor_2_dB>,<Delta_power>,<Delta_time>[,<Average>]<Delta_power> and

Página 35 - OPERATION SENSOR MENU

Chapter 2Installation2-1INTRODUCTIONThis chapter provides information for the initial inspection and preparationfor use of the ML2430A Series Power Me

Página 36

Syntax:GRPIX <mode>modeNORMMINMAXMINMAXAVGRemarks:Changes the type of graph displayed:NORM: Profiles the sensor readings vs. time from the trigg

Página 37

Remarks:Sets the pre trigger percentage of the profile screen. The percentage ofthe data collection period that shows pretrigger infomation if the dis

Página 38

RelatedCommands:GRSWPGRSWRQuery:GRSWS?ReturnedString:GRSWS <state>GRTMM Profile Min/Max tracking modePROFILE SETUPSyntax:GRTMM <mode>mode:

Página 39

val:–150.0 to +250.0Remarks:It is not necessary to specify units as the displayed units are always as-sumed. Profile and Source Sweep modes always use

Página 40

Remarks:Sets the profile trigger arming ON or OFF. If set to ON, the system firstchecks to see if the BNC sweep blanking input is TRUE before it start

Página 41 - OPERATION CHANNEL MENU

Syntax:GTLVL <val>val:–30 to +20 dBmRemarks:When the system trigger in profile mode is set to either INTA or INTB (in-ternal sensor A or B) it w

Página 42 - CHANNEL MENU OPERATION

GTTYP Set profile trigger type TRIGGERSyntax:GTTYP <type>type:RISEFALLRemarks:When the profile system trigger source is set to INTA or INTB (Int

Página 43

val:Remarks:Sets the high limit. The HLIMS command turns the limits on and off. It isnot necessary to enter the units as the limit value is checked ag

Página 44 - TRIGGER MENU OPERATION

Syntax:HOLD <state>state:ON or OFFRemarks:This command holds the present graph displayed on the screen and isavailable in all graph modes. In Pr

Página 45 - OPERATION TRIGGER MENU

state:ON or OFFRemarks:This will link the trigger set-up between Profile mode and Readout modeso that the sample delay and the gate width will agree.

Página 46

a fuse mounted inside the unit, on the main board. Agrounding terminal is provided on the rear panel toground the unit during operation from a DC supp

Página 47

Remarks:The LLIMS command turns the limits on and off.RelatedCommands:LLIMQuery:LLIMS? <c>ReturnedString:LLIMS <c>,<state>MMRST Min

Página 48

MNGD <number_of_elements>,<element_1>,<element_2>,<ele-ment_n>...<\n>The first number in the string is the number of ele

Página 49 - OPERATION SYSTEM MENU

Syntax:MODLIM <state>value:TRUE or FALSERemarks:When set to TRUE, produces an SRQ and autodials the phone number(set with MODPH) when a channel

Página 50 - SYSTEM MENU OPERATION

Syntax:MODRED <count>count:0 to 10Remarks:Sets the number of retrys after a failure to connect. The delay betweenretrys is set using MODDEL. See

Página 51

Syntax:MXGDRemarks:Available in graph modes only. Outputs in ASCII form the max graph data.The format is as follows:MXGD <number_of_elements>,&l

Página 52

port:1 or 2c:1 or 2Remarks:This command changes the channel represented by BNC output modesthat can take data from either channel 1 or 2, such as “Ana

Página 53

Query:OBDST? <port>ReturnedString:OBDST <port>,<units>,<val>OBMD BNC output mode selectBNCSyntax:OBMD <port>,<mode>

Página 54

port:1 or 2val:–5.00 to +5.00 VoltsRemarks:Sets up the stop value for the voltage output in analog output mode. At-tempting to set the start value to

Página 55

OFFFIX Offset fixed value SENSORSyntax:OFFFIX <s>,<val>[units]s:A or Bval:–99.999 to +99.999units:dBRemarks:The value added to the sensor

Página 56

ReturnedString:OFFTBL <s>,<val>OFFTBR Output an offset tableSENSORSyntax:OFFTBR <val>val:1 to 5Remarks:Outputs the selected offset t

Página 57

FusesThe ML2430A Series Power Meter AC and DC inputlines are protected by internally mounted fuses. Thesefuses should only be changed by qualified ser

Página 58

–10.234 becomes 10479, converted to hexadecimal FFFFD711. See theprogramming examples for more detail.OFFTYP Offset type to useSENSORSyntax:OFFTYP <

Página 59

For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divideby 1024 to get the dB value (–11.652).OGD Output Graph DataDATA OUTPUTSyntax:OG

Página 60

Syntax:ON <c>,<val>c:1 or 2val:1 to 1000Remarks:Readout and Power vs. Time modes only. This command returns the speci-fied number of readi

Página 61

and measurement channel setups.When in Readout mode, the Channel 1 and Channel 2 values, and themax/min values if present, are printed below the heade

Página 62 - CAL/ZERO MENU OPERATION

range 3 upper, range 3 lower, range 4 upper, range 4 lower, range 5 up-per, range 5 lower.REL Relative controlCHANNELSyntax:REL <c>,<mode>

Página 63 - OPERATION CAL/ZERO MENU

Remarks:This function is used to toggle a sensor’s range hold off or on, to set a spe-cific range to be held, or to select AUTO ranging. RGH sent with

Página 64 - Procedures

Remarks:Normally when the system is powered on the ML2430A Series returns tothe state it was in when it was powered off. This includes all the offset

Página 65 - SENSOR CALIBRATION PROCEDURES

SENSTL Set Sensor Settle Percentage SENSORSyntax:SENSTL <s>,<val>s:A or Bval:0.01 to 10%Remarks:Sets how long the system waits for the sig

Página 66 - PROCEDURES SENSOR ZERO/CAL

Remarks:Determines the stop frequency when in frequency sweep mode.Query:SRCSPFRQ?ReturnedString:SRCSPFRQ <frequency>SRCSPPWR Source Sweep Stop

Página 67 - PRINTER CONNECTION PROCEDURES

powervalue:powerRemarks:Determines the start power level of power sweep mode.Query:SRCSTPWR?ReturnedString:SRCSTPWR <power>START Initial startup

Página 68 - ML2430A OM 5-5

The required parts and tools are listed below:Tools Required: Small Phillips screw driverLarge Phillips screw driverSmall Phillips torque screw driver

Página 69 - 5-6 ML2430A OM

B–A, '5' = A/B, '6' = B/A, '7' = EXT Volts.D = Channel 1 units: '0' = dBm, '1' = Watts, '2&apos

Página 70 - ML2430A OM 5-7

V = Sensor A zeroed status: '0' = Not zeroed, '1' = Zeroed.W = Sensor B Zeroed status: '0' = Not zeroed, '1' =

Página 71 - 5-8 ML2430A OM

Remarks:Sets the passed store number to the setup contained in the binary datathat was extracted using the SYSRD command.RelatedCommands:SYSRDSYSLNM S

Página 72 - ML2430A OM 5-9

RelatedCommands:SYSLDTEXT User text commandSYSTEMSyntax:TEXT <text string>text string:Text string of up to 20 charactersRemarks:Defines the text

Página 73 - 5-10 ML2430A OM

RelatedCommands:TR1, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2TR1 Trigger immediateGPIB TRIGGERSyntax:TR1 <c>c:1 or 2Remarks:Tr

Página 74 - ML2430A OM 5-11

After a TR2 command the instrument returns to either TR0 (trigger hold) orTR3 (trigger free run) mode depending on what it was previously set to.Relat

Página 75 - 5-12 ML2430A OM

c:1, 2 or 1&2val:0.0 to 1.0 secondsRemarks:The time the system waits after a trigger event has happenedbefore taking measurements when in READOUT

Página 76 - ML2430A OM 5-13

Remarks:If the Trigger source is set to INTA or INTB (internal A or B) the systemtriggers on a rising or falling power level edge. Use this command to

Página 77 - 5-14 ML2430A OM

MANUAL (manual push button trigger)CONT (continuous)Remarks:This command is overridden by the TR0, TR1 and TR2 commands whenin READOUT or POWER vs. TI

Página 78 - ML2430A OM 5-15

type:RISEFALLRemarks:Sets the control type of the external trigger input used when the triggersource is set to EXTTTL in READOUT or POWER vs. TIME mod

Página 79 - SOURCE SWEEP MODE PROCEDURES

Locate the support bracket on the four PM case pillars.Secure with 4 screws905-68 and 4 washers 900-345.(See max.torque settings page 2-4.)5.Locate th

Página 80 - PROCEDURES SOURCE SWEEP MODE

6-11GPIB EMULATIONMODESThe ML2430 Anritsu power meter emulates the GPIB communication of otherpower meters. The emulation mode can be set through the

Página 81 - TIME MODE PROCEDURES

6-12ML4803AEMULATIONCOMMANDSThis section provides an alphabetical listing of the GPIB commands (mnemonics)used to program the Model ML2430A Series Pow

Página 82 - PROCEDURES USER CAL FACTORS

Output Requests There are three commands to request output from the ML4803A: OPW for areading, ODT for the cal factor, offset and reference values, an

Página 83 - USER CAL FACTORS PROCEDURES

CCA Clear the calfactor value to zero.CDJ Perform a CAL 0 dBm.Remarks:During the cal 0 dBm sequence, the CAL execution bit in the status byte isset. W

Página 84 - ML2430A OM 5-21

When selecting GPIB emulation modes, the instrument configures itself tothe preset conditions of the instrument to be emulated. For example, whenselec

Página 85 - 5-22 ML2430A OM

MCR Clears the reference value at the specified memory loca-tion.Syntax:MCR<mem>mem:Memory location 1 to 30.Remarks:Clears the reference value a

Página 86 - ML2430A OM 5-23

Set the offset value at the specified memory location indBm.Syntax:MOF<mem><;><value>mem:Memory location 1 to 30.value:Offset value

Página 87 - 5-24 ML2430A OM

OFF Set sensor offset valueSyntax:OFF<value>value:Offset value to add to the sensor reading.OI? Request identitySyntax:OI?Remarks:Response: <

Página 88 - GPIB Operation

OPW Request for channel reading.Remarks:Outputs measuring condition, measured data, and status. CR and LFcodes are automatically output and executed a

Página 89 - 6-2 ML2430A OM

Output Code Contents FunctionARG1ARG2ARG3ARG4ARG5AUTOhighest sensitivity 1234lowest sensitivity 5RANGEAVE0AVE9AVE1AVE2AVE3AVE4OFFHOLD1 (1 second inter

Página 90 - ML2430A OM 6-3

ML2400A-03 RackMount InstallationThis section describes the assembly procedure for fitting two ML2430A PowerMeters (PM) into a instrument rack. The PM

Página 91 - QUERY COMMANDS GPIB OPERATION

As shown in the examples above, the dBm data is output in fixed rotation,while the WATT data is output in scientific notation. The exponent may beconv

Página 92 - ML2430A OM 6-5

Remarks:When SRQ0 is issued, the SRQ will no long turn off and on with eachreading. The SRQ is set back on by the SRQ1 command or by requestingdata.ST

Página 93 - 6-6 ML2430A OM

6-13HP 436AEMULATIONCOMMANDSThis section provides an alphabetical listing of the commands (mnemon-ics) used to program the Model ML2430A Series Power

Página 94 - ML2430A OM 6-7

dB (rel)Remarks:Set to dB units in relative mode using the present relative reference value.C dB (ref)Remarks:Set to dB units in relative mode using t

Página 95 - 6-8 ML2430A OM

I Trigger without settling.Remarks:Triggers a single reading which is added to the internal digital filter and theupdated filter power level is return

Página 96

Output Format The output data format for the HP 436A emulation mode is shown below.Table 6-2 (next page) describes the HPIB output data format.6-98 ML

Página 97 - GROUPS GPIB OPERATION

ML2430A OM 6-99GPIB OPERATION HP 436A EMULATION COMMANDSDefinition CharacterSTATUS ASCII DecimalMeasured value validWatts mode under rangeOver rangeUn

Página 98 - GPIB OPERATION FUNCTIONAL

6-14HP 437B EMULATIONCOMMANDSThis section provides an alphabetical listing of the commands (mnemon-ics) used to program the Model ML2430A Series Power

Página 99 - 6-12 ML2430A OM

*ESE? Return Event status register enable maskSyntax:*ESE?Remarks:Returned format: <unsigned character>When converted to an 8-bit binary number,

Página 100 - ML2430A OM 6-13

*RST Reset DeviceSyntax:*RSTRemarks:Resets the ML2430A Series to the default configuration (see Appendix A,Section A-3, or see the HP manual when in H

Página 101 - 6-14 ML2430A OM

ToolsRequired: 1 Small Phillips screw driver1 Large Phillips screw driver1 Small Phillips torque screw driver 10cNm to 120cNm.1 Assembly drawing “ML24

Página 102 - ML2430A OM 6-15

Remarks:Performs a self test and returns 000.'RelatedCommands:STERR@1 Set SRE maskSyntax:@1<val>val:8-bit maskRemarks:Status Byte Structure

Página 103 - 6-16 ML2430A OM

Remarks:Same as the *CLS command. Resets all of the GPIB status registers andclears the input queue.CT Clear the cal factor tableSyntax:CT<table_nu

Página 104 - ML2430A OM 6-17

DE display enableSyntax:DERemarks:Return the display to normal operation after the display has been set in DDmode.RelatedCommands:DD, DFDF Display dis

Página 105 - 6-18 ML2430A OM

HP438A (Hewlett-Packard)ML4803 (Anritsu ML4803A Series)Remarks:Set the GPIB emulation to emulate other types of power meters. This com-mand is availab

Página 106 - ML2430A OM 6-19

FA Auto averageSyntax:FARemarks:Automatic Filter on. Allows the system to automatically select the filterused to reduce the jitter in the display.Rela

Página 107 - 6-20 ML2430A OM

Example:To set the frequency of the input signal to 300 MHz:FR300MZGT Set group triggerSyntax:GT<mode>mode:012Remarks:0 = Ignore Group Execute T

Página 108 - ML2430A OM 6-21

terminator:%PCTENRemarks:The calibration factor compensates for mismatch losses and effective effi-ciency over the frequency range of the power sensor

Página 109 - 6-22 ML2430A OM

LN Set linear unitsSyntax:LNRemarks:Changes the display to linear units (Watts).OC Set calibrator stateSyntax:OC<state>state:0 (OFF)1 (ON)Remark

Página 110 - ML2430A OM 6-23

state:0 (OFF)1 (ON)Remarks:For example: OF1 (Turn offsets ON).OI Return identification stringSyntax:OIRemarks:Returned format:<company name>,<

Página 111 - 6-24 ML2430A OM

val:1 to 10Remarks:The ML2430A Series can store up to 10 instrument configurations for con-venient recall. The configuration parameters stored are the

Página 112 - ML2430A OM 6-25

WARRANTYThe ANRITSU product(s) listed on the title page is (are) warranted against defects in materials andworkmanship for one year from the date of s

Página 113 - 6-26 ML2430A OM

The procedure for fitting a ML2430A to a HP34401A Multimeter is as fol-lows:1. Fit to the ML2430A Power Meter front brackets C37276, C37275 andrear br

Página 114 - ML2430A OM 6-27

RL Relative modeSyntax:RL<mode>mode:012Remarks:Relative mode permits any measurement result to be compared in dB orpercent to a reference value.

Página 115 - 6-28 ML2430A OM

Remarks:Selects the cal factor table to be used.SM status messageSyntax:SMRemarks:Returns the status message in the format:AAaaBBCCccDDddEFGHIJKLMNOP&

Página 116 - ML2430A OM 6-29

Remarks:Stores the present configuration to the selected register.Example:To store the current instrument configuration in register 2:ST2ENRelatedComm

Página 117 - 6-30 ML2430A OM

Remarks:Triggers a new series of readings; enough to update the internal digital fil-ter for a noise free reading at the current power level. The valu

Página 118 - ML2430A OM 6-31

6-15HP 438AEMULATIONCOMMANDSThis section provides an alphabetical listing of the GPIB commands (mne-monics) used to program the Model ML2430A Series P

Página 119 - 6-32 ML2430A OM

Remarks:Display the Input A reading minus the Input B reading.AP Set single sensor A displaySyntax:APRemarks:Set the display to output Input A reading

Página 120 - ML2430A OM 6-33

terminator:%PCTENRemarks:Same as the ML24XXA (native) CFADJ command. Sets a calibration factorto be used when performing a 0 dBm calibration.Examples:

Página 121 - 6-34 ML2430A OM

RelatedCommands:DDDR Set GPIB addressSyntax:DR<val>val:1 to 30Remarks:Changes the device address when operating in HP emulation mode. Thepower m

Página 122 - ML2430A OM 6-35

Remarks:Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-ter setting. This function is the same as the AVGM command.Rela

Página 123 - 6-36 ML2430A OM

Remarks:The calibration factor compensates for mismatch losses and effective effi-ciency over the frequency range of the power sensor.Examples:KB99.9%

Página 124 - ML2430A OM 6-37

2-7BATTERYCHARGING,REMOVAL ANDREPLACEMENTThe optional ML2430A Series Power Meter battery is a 12 Volt, 3000mA-h nickel-metal hydride (Ni-MH) multi-cel

Página 125 - 6-38 ML2430A OM

LN Set linear unitsSyntax:LNRemarks:Changes the display to linear units (Watts).OC Set calibrator stateSyntax:OC<state>state:0 (OFF)1 (ON)Remark

Página 126 - ML2430A OM 6-39

Reference Oscillator = OffActive entry channel = AMeasurement units = WattsREL mode = offMeasurement parameters (set for Sensor A and Sensor B):Cal Fa

Página 127 - 6-40 ML2430A OM

mode:012Remarks:Relative mode permits any measurement result to be compared in dB orpercent to a reference value. Relative mode can be enabled using t

Página 128 - ML2430A OM 6-41

CC: sensor A rangecc: sensor B rangeDD: sensor A filterdd: sensor B filterE: measurement unitsF: active entry channelG: OSC statusH: REL mode statusI:

Página 129 - 6-42 ML2430A OM

Remarks:Triggers a single reading which is added to the internal digital filter and theupdated filter power level is returned on the GPIB. After a TR1

Página 130 - ML2430A OM 6-43

6-16PROGRAMMINGEXAMPLESThe following programming examples are provided as a general guideline on howto program the ML2430A Series Power Meters using G

Página 131 - 6-44 ML2430A OM

Get Graph DataFunction GetGraphData ()‘function assumes that you have the graph display‘setup and that there is a global array called‘Graph_Data().‘ma

Página 132 - ML2430A OM 6-45

StatusRegisterControlThis function demonstrates how to use the Status Registers to provide synchroni-zation.Uses the TR2 (trigger with settling) comma

Página 133 - 6-46 ML2430A OM

488.2General Send/Receive SystemThis function uses the status registers to synchronize the GPIB commands andreturn data if a query command was used. T

Página 134 - ML2430A OM 6-47

4882SendReceive= Execution_ErrorElseIf (Val(ESB) And DEVICE_ERR_bit) Then4882SendReceive = Device_dependent_ErrorEnd IfEnd Function6-132 ML2430A OMPRO

Página 135 - 6-48 ML2430A OM

For optimal battery life, the battery should be fully discharged before re-charging. Repeated partial charge/discharge cycles can result in a loss ofb

Página 136 - ML2430A OM 6-49

Binary OutputMessage DecodingThe following program example may be used to decode the three types of binaryoutput messages. The commands that reference

Página 137 - 6-50 ML2430A OM

/* Read and cast the data */while (count < length){real_data1[loop++] = ((float)(*bin_value++))/1024.0;count += 4;}real1_entries = loop;}else if (d

Página 138 - ML2430A OM 6-51

else if (decode_type == CAL_FACTOR_BINARY_DATA){/* Decode header *//* Read length of binary data*/ch_val[0] = buffer[6];ch_val[1] = buffer[7];ch_val[2

Página 139 - 6-52 ML2430A OM

Appendix ASpecificationsA-1INTRODUCTIONThis appendix provides system specifications for the ML2430A SeriesPower Meters along with listings of system d

Página 140 - ML2430A OM 6-53

A-2 ML2430A OMSYSTEM SPECIFICATIONS APPENDIX AInstrument Ac-curacy:< 0.5 %Zero Set andDrift:< 0.5 % MA2420A; < 0.5 % MA2470A Series andMA2440

Página 141 - 6-54 ML2430A OM

ML2430A OM A-3APPENDIX A SYSTEM SPECIFICATIONSCursors:Two manually adjustable cursors with power, delta cur-sor power, between cursor power average, a

Página 142 - ML2430A OM 6-55

A-4 ML2430AOMSYSTEMSPECIFICATIONS APPENDIXACal Factor Volt-ageInput(BNC):Operating Modes:Voltage:Display voltage reading on selected channel.Voltage p

Página 143 - 6-56 ML2430A OM

APPENDIX A SYSTEM SPECIFICATIONSDisplay:Flat panel monochrome LCD graphic with backlight.Operating Tem-peratureRange:0 to 50° C (32 to 122° F)StorageT

Página 144 - ML2430A OM 6-57

A-3SYSTEM DEFAULTSThe following default parameters are loaded whenever preset is selectedfrom the front panel or through GPIB.A-6 ML2430A OMSYSTEM DEF

Página 145 - 6-58 ML2430A OM

ML2430A OM A-7APPENDIX A SYSTEM DEFAULTSfrequencies 0 Hzoffset values 0 dBSENSOR - duty cycleduty cycle state offduty cycle 100%CHANNEL - setupinput c

Página 146 - ML2430A OM 6-59

2-8STORAGE ANDSHIPMENTThe following paragraphs describe preparing the power meter for storage andshipment.Preparation forStoragePreparing the power me

Página 147 - 6-60 ML2430A OM

A-8 ML2430A OMSYSTEM DEFAULTS APPENDIX ASYSTEM - setup = profileTRIGGER - setupsource continuoussample delay 1 mssample gate width 20 mssource = conti

Página 148 - ML2430A OM 6-61

ML2430A OM A-9APPENDIX A SYSTEM DEFAULTSSYSTEM - displaybattery backlight ondisplay contrast 5peak meter display offfrequency/offset display offGPIB u

Página 149 - 6-62 ML2430A OM

A-10 ML2430A OMSYSTEM DEFAULTS APPENDIX Aoutput 2 onlymode = RF blankingchannel 1output TTL during zero highinput 1blanking active TTL level highSYSTE

Página 150 - ML2430A OM 6-63

A-4SYSTEM ERRORMESSAGESThis section details some of the front panel error messages that may ap-pear. In most cases, the error condition can be easily

Página 151 - 6-64 ML2430A OM

Error Message MeaningSENSOR A not fittedREADING SENSOR A dataSensor A is required for the currentmeasurement, but is not fitted.as per messageSENSOR B

Página 152 - ML2430A OM 6-65

Appendix BGPIB Quick ReferenceB-1INTRODUCTIONThis appendix contains quick reference tables for all supported com-mands. Native mode commands are sorte

Página 153 - 6-66 ML2430A OM

Table B-3. Channel Functional Group CommandsCommand Function PageCHCFG(?) Channel input configuration 6-28CHRES(?) Set channel decimal point resolutio

Página 154 - ML2430A OM 6-67

Table B-6. GPIB 488.2 Functional Group CommandsCommand Function Page*CLS Clear GPIB status bytes 6-12*ESE Event Status Byte enable 6-7, 6-12,6-100*ESE

Página 155 - 6-68 ML2430A OM

Table B-9. Profile Setup Functional Group CommandsCommand Function PageCURLK(?) Link cursors in profile mode 6-30DTRGD(?) Display Trigger Delay 6-34GR

Página 156 - ML2430A OM 6-69

Command Function PageOFFCLR Clear an offset table 6-60OFFFIX(?) Offset fixed value 6-61OFFTBL(?) Select an offset table 6-61OFFTBR Output an offset ta

Página 157 - 6-70 ML2430A OM

STORAGE AND SHIPMENT INSTALLATION2-12/2-12 ML2430AUNITED STATESANRITSU SALES COMPANY685 Jarvis DriveMorgan Hill, CA 95037-2809Telephone: (408) 776-830

Página 158 - ML2430A OM 6-71

SRCSTAT Source sweep status request 6-71SYSLNM Saved set naming 6-75TEXT(?) User text command 6-76TEXTS(?) User text display command 6-76Table B-12. T

Página 159 - 6-72 ML2430A OM

Query Returned String PageDBLTIM? DBLTIM <VAL> 6-32DCONT? DCONT <VAL> 6-32DISP? DISP <STATE> 6-33DPEAK? DPEAK <MODE> 6-33DTRGD

Página 160 - ML2430A OM 6-73

Query Returned String PageOBDSP? <PORT> OBDSP <PORT>,<UNITS>,<VAL> 6-45OBDST? <PORT> OBDST <PORT>,<UNITS>,&l

Página 161 - 6-74 ML2430A OM

B-3ML4803A QUICKREFERENCEThe following tables reference ML4803A emulation mode commands.Table B-14. ML4803A Emulation Mode CommandsCommand Function Pa

Página 162 - ML2430A OM 6-75

B-4HP 436A QUICKREFERENCEThe following table references HP 436A emulation mode commands.Table B-15. HP 436A Emulation Mode CommandsCommand Function Pa

Página 163 - 6-76 ML2430A OM

B-5HP 437B QUICKREFERENCEThe following table references HP 437B emulation mode commands.Table B-16. HP 437B Emulation CommandsCommand Function Page*CL

Página 164 - ML2430A OM 6-77

Command Function PageGT Set group trigger 6-108,6-121ID Return identification string 6-108IDN? HP437 identity request 6-108KB Calibration factor 6-108

Página 165 - 6-78 ML2430A OM

Command Function PageTR1 Trigger immediate 6-77,6-115,6-126TR2 Trigger with a settling delay 6-77,6-115,6-127TR3 Trigger free run 6-78,6-116,6-127ZE Z

Página 166 - ML2430A OM 6-79

B-6HP 438A QUICKREFERENCEThe following table references HP 438A emulation mode commands.Table B-17. HP 438A Emulation CommandsCommand Function Page?ID

Página 167 - 6-80 ML2430A OM

Command Function PageOI Identification 6-64, 6-97,6-111,6-123OS Set offset value 6-111,6-123PR Preset the unit 6-111,6-123RA Auto Range 6-111,6-124RC

Página 168 - ML2430A OM 6-81

Chapter 3Connections3-1INTRODUCTIONThis chapter describes physical connections to the power meter on boththe front and rear panels.3-2FRONT PANELCONNE

Página 169 - 6-82 ML2430A OM

B-7HP-IB SUPPORTThe following tables list HP-IB commands for the HP 437B and HP 438Apower meters, and which commands are supported in the AnritsuML243

Página 170 - ML2430A OM 6-83

Mnemonic Action Supported? RestrictionsLM1 Limit check on Yes Both high and low on as HP 437BLN Units to Watts Yes NoneLP Learn mode 1 No Not supporte

Página 171 - 6-84 ML2430A OM

HP 438A CommandsMnemonic Action Supported? RestrictionsAD Config to A-B Yes NoneAE Select sensor A Yes NoneAP Config to A Yes NoneAR Config A/B Yes No

Página 172 - 76543210

Mnemonic Action Supported? RestrictionsRV Read SRE Yes NoneSM Status ouput Yes As much as has meaning for ML2430A.Set to 0 if not usedST Store setup Y

Página 173 - 6-86 ML2430A OM

Appendix CMenu MapsC-1INTRODUCTIONThe ML2430A Series Power Meter is driven by five main menus; Sensor, Chan-nel, Trigger, System, and Cal/Zero, each a

Página 174 - ML2430A OM 6-87

[CFSRC, CFVAL] SOURCEFrequency | manual | vghzSETUP } if source = VGHz only[CVSTF] START Freq10kHz to 122GHz[CVSPF] STOP Freq10kHz to 122GHz[CVSTV] ST

Página 175 - 6-88 ML2430A OM

10kHz to 122GHzdetermined by fitted sensorFACTOR150% to 0.07%–1.76dB to 31.55dBDONEDELETE[CFUSAV] SAVE to sensorNOYES[CFUCT] CLEAR tableNOYES[CFUPT] P

Página 176 - ML2430A OM 6-89

A|BAveraging MODEAuto | moving | repeat | off[AVG] Averaging NUMBER } if mode = MOVING or REPEAT1to512[AVGLL] Auto LOW LEVEL averagingOff \ low \ medi

Página 177 - 6-90 ML2430A OM

upENTRY1to200[OFFCLR] CLEAR selected table } if type = TABLEDuty cycleSENSOR } not in ML2437AA|B[DUTYS] Duty cycle STATEOn | off[DUTY] DUTY cycle100%

Página 178 - ML2430A OM 6-91

C-3CHANNEL MENUThe Channel menu controls the operation of a display channel. There are twodisplay channels, Channel 1 and Channel 2. Channel 1 appears

Página 179 - 6-92 ML2430A OM

An optional rear panel Calibrator connector is offered as an alternative (see Fig-ure 3-2). If the rear panel connector option is installed, the front

Página 180 - ML2430A OM 6-93

[HLIMS] HIGH StateOn | off[LLIMS] LOW StateOn | off[FHOLD] Fail indicator HOLDOn | off[FBEEP] Fail BEEP controlOn | offML2430A OM C-7APPENDIX C CHANNE

Página 181 - 6-94 ML2430A OM

C-4TRIGGER MENUThe Trigger menus are always available in PROFILE operation mode, as selectedfrom the System menu. In READOUT and POWER vs. TIME modes,

Página 182 - ML2430A OM 6-95

C-5SYSTEM MENUThe System menus control the operating modes, display visibility, sound, rearpanel functions, and battery state of the ML2430A Series Po

Página 183 - 6-96 ML2430A OM

PwrVsTime } if system setup mode = Power vs. Time[GRMD] CHANNEL1|2[GRPIX] DATA HOLD representationNormal | average | min&max | min | max[GRDDT] Da

Página 184 - ML2430A OM 6-97

[GRDATA, GRDRQ] READOUT[GPRST] CLEAR[CURLK] LINK CURSOR[HOLD] Graph HOLD[GRAUTO] AUTO scaleDisplay[DBLGHT] Battery BACKLIGHTOn | timed | off[DCONTD, D

Página 185 - 6-98 ML2430A OM

Battery[BAUTS] AUTO power off } if smart battery detectedOn | off[BAUTT] Auto power off TIME } if smart battery detected10 to 240 minutesSTATUS } if s

Página 186 - ML2430A OM 6-99

[MODPWR] POWER on/offTrue | falseBNC setupPORTOutput 1, output 2, input 1, input 2[OBMD] Operating MODE only if port = output 1 or output 2o/p 1off |

Página 187 - 6-100 ML2430A OM

if mode = RF BLANKING and port = output 2[OBCH] CHANNEL1|2[OBZL] Output TTL during zeroingLow | highif port = input 1[IBBLP] Blanking active TTL LEVEL

Página 188 - Status Byte Register

C-6CAL/ZERO MENUThe Cal/Zero menu establishes the 0.0 dBm reference calibration and zeroing ofthe sensors. Refer to Chapter 5 for specific procedures.

Página 189 - 6-102 ML2430A OM

AAC . . . . . . . . . . . . . . . . . . . . . 2-1, 3-3, 4-2accessoriespower meter . . . . . . . . . . . . . . . .

Página 190 - ML2430A OM 6-103

ML2430A OM 3-3/3-4OPERATION REAR PANEL CONNECTORSFigure 3-2. ML2430A Series Rear PanelGPIB/ IEEE488 ConnectorStandard General Purpose Inter-face Bus c

Página 191 - 6-104 ML2430A OM

controls . . . . . . . . . . . . . . . . . . 4-1 - 4-2operation . . . . . . . . . . . . . . . . . . . . 4-1functio

Página 192 - ML2430A OM 6-105

Rrack mount . . . . . . . . . . . . . . . . . . 1-2 - 2-7RAM . . . . . . . . . . . . . . . . . . . . . . . . 4

Página 194 - ML2430A OM 6-107

Chapter 4Front Panel Operation4-1INTRODUCTIONThe ML2430A Series Power Meter is controlled from the front panel using thefive main menu keys; Sensor, C

Página 195 - 6-108 ML2430A OM

Safety SymbolsTo prevent the risk of personal injury or loss related to equipment malfunc-tion, ANRITSU Company uses the following symbols to indicate

Página 196 - ML2430A OM 6-109

4-2 ML2430A OMFRONT PANEL CONTROLS OPERATIONFigure 4-1. ML2430A Series Front Panel ControlsNumeric KeypadProvides the means for enteringnumeric data i

Página 197 - 6-110 ML2430A OM

4-3POWER-ONPROCEDUREAt power-on, the power meter performs a brief power-on self test (POST).After the POST, the instrument loads the last used configu

Página 198 - ML2430A OM 6-111

If any error, other than a DSP error, is encountered, the text:“Press ANY key to continue”will be displayed at the bottom of the screen. If only warni

Página 199 - 6-112 ML2430A OM

4-4SENSOR MENUThe Sensor menu has controls for sensor data processing.NOTEWhen editing an entry in a menu, pressing theCLRkeyclears the digits.If in a

Página 200 - ML2430A OM 6-113

HOLD[RGH]Allows the operating range of the selected sensor to be set to thedesired range. Select a Range Hold value of 1 to 5, or Auto. Whenin Auto, t

Página 201 - 6-114 ML2430A OM

For greater accuracy, calibration factors are interpolated for settingsthat are between the calibration factor data provided in the sensorEEPROM. For

Página 202 - ML2430A OM 6-115

FactorySelects the Factory calibration factor table. Pressing Factory andthe +/- key on the numeric keypad allows selection of a user-defined table in

Página 203 - 6-116 ML2430A OM

SETUP[CVSTFCVSPFCVSTVCVSPV]Sets up the Start and Stop frequencies and voltages when Sourceis set to V/GHz. This tells the ML2430A how to determine the

Página 204 - ML2430A OM 6-117

REPEAT averaging also allows the user to manually select theamount of averaging regardless of the signal level, however the dis-play is updated only w

Página 205 - 6-118 ML2430A OM

RESETSweep average reset. If the graph sweep averaging state is ON,this key resets the data points and restarts the sweep to sweepmode.CURSOR[GRSWR]Be

Página 206 - ML2430A OM 6-119

SAFETY-2 ML2430A OMWhen supplying AC power to this equipment, connect the accessory 3-pinpower cord to a 3-pin grounded power outlet. If a grounded 3-

Página 207 - 6-120 ML2430A OM

TABLE[OFFTBLOFFTBROFFTBUOFFVAL]Select the offset table number (1-5) when Offset TYPE is set to Ta-ble. When a table is selected, additional soft keys

Página 208 - ML2430A OM 6-121

4-5CHANNEL MENUThe Channel menu controls the operation of a display channel. There are twodisplay channels, Channel 1 and Channel 2. Channel 1 appears

Página 209 - 6-122 ML2430A OM

RESET[MMRST]This function resets the Min/Max (when ON) for the channelselected.Rel 1[REL]After the relative power level is set by the operator, the Re

Página 210 - ML2430A OM 6-123

have been set for –10 dBm and the display units are subsequentlychanged from dBm to Watts, the system still checks for the readingto rise above –10, e

Página 211 - 6-124 ML2430A OM

4-6TRIGGER MENUThe Trigger function in the ML2430A allows the user to define under what condi-tions measurements are taken, and the time period they a

Página 212 - ML2430A OM 6-125

The trigger icons appear as shown in Figure 4-2.Only when a channel input configuration includes a sensor with a measurementmode that requires an icon

Página 213 - 6-126 ML2430A OM

DELAY[TRGDLYGTDLY]In Profile mode, DELAY sets the time delay (after the display triggerdelay) to when the system starts to take and display readings,

Página 214 - ML2430A OM 6-127

1. Connect to the rear panel digital input.2. Select Trigger|Setup|ARMING|Blanking ON.3. Set the polarity of the blanking (System menu)Example power m

Página 215 - 6-128 ML2430A OM

Figure 4-4 shows a typical trigger timing diagram. Note that the displaytrigger delay is only present when in Profile operation mode, and helps insett

Página 216 - ML2430A OM 6-129

4-7SYSTEM MENUThe System menus control the operating modes, display visibility, sound, rearpanel functions, and battery state of the ML2430A Series Po

Página 218 - ML2430A OM 6-131

With LINK readout/profile trigger set to ON, switching to Readoutmode uses the same trigger conditions, but allows the full dynamicrange of the meter

Página 219 - 6-132 ML2430A OM

DATA HOLD[GRPIX]This selects how the graph is displayed on the screen. Select fromNORMAL, Min&Max, Min, or Max.NOTEIf either Min&Max, Min, or

Página 220 - ML2430A OM 6-133

If the source used does not provide a blanking output, the blanking signal may bedisabled as follows: select System|Rear Panel|BNC. Select PORT until

Página 221 - 6-134 ML2430A OM

STOP[SRCSPFRQSRCSPPWR]Sweep stop frequency (MHz or GHz) or power (dBm)NOTEWhen the power meter is communicating with a signalsource/generator over the

Página 222 - ML2430A OM 6-135

1 cursor 1 reading2 cursor 2 readingD P Power difference between cursor 1 and cursor 2D t Time difference between cursor 1 and cursor 2AV Between curs

Página 223 - Specifications

2 cursor 2 readingD P Power difference between cursor 1 and cursor 2X1 X axis at cursor 1X2 X axis at cursor 2Data hold = MIN (or MAX) :1 cursor 1 min

Página 224 - A-2 ML2430A OM

ContrastDOWN[DCONTDDCONT]Reduces the display contrast. Adjust to suit ambient conditions.ContrastUP[DCONTUDCONT]Increases the display contrast. Adjust

Página 225 - ML2430A OM A-3

Print[PRINT]This selection prints the screen and various operational settings through the rearpanel printer port.BatteryControls battery setup when th

Página 226 - A-4 ML2430AOM

NOTEImmediately after power-on, the “estimated operating timeremaining” displayed may not be genuine, as the batteryrequires a few minutes to calculat

Página 227 - ML2430A OM A-5

Source IF allows the power meter to communicate with an Anritsu68/69000-series synthesizer when the operation mode is set toSource Sweep.BAUD[RSBAUD]S

Página 228 - SYSTEM DEFAULTS APPENDIX A

Table of ContentsChapter 1 - General Information1-1 Scope of This Manual . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 229 - APPENDIX A SYSTEM DEFAULTS

POWER [MODPWR] – If this is set, and the power cycles on themeter, then the number specified in the “phone number” field will bedialed. When a connect

Página 230

ACMod output (port 1 only) is a TTL signal synchronized to the in-ternal chopper (when used) of the signal channel. This signal canbe used for synchro

Página 231

Graphics This menu presents additional graphic display controls:CONNECT[GRCP]This control is normally ON and causes the data between samplesto be inte

Página 232

RF ON/OFF[RFCAL]Turns the RF calibrator ON or OFF.Ext V[VZERO]Zeros the rear panel multi-purpose BNC connector used for Volts per GHz con-nection (Ana

Página 233 - ML2430A OM A-11

Chapter 5Procedures5-1INTRODUCTIONThis chapter presents some common procedures for use with theML2430A Series Power Meter. These procedures refer to t

Página 234 - A-12 ML2430A OM

To zero the sensor without calibration, press the Cal/Zero front panel keyand the Zero soft key, then select the appropriate sensor.Note that if only

Página 235 - GPIB Quick Reference

Any error conditions encountered during calibration, for example the pres-ence of extraneous noise or RF signals, will result in an error message onth

Página 236 - B-2 ML2430A OM

5-7PRINTERCONNECTIONSee Chapter 3, Connectors, for the location of the parallel port connector on therear panel. Connect a parallel printer cable from

Página 237 - ML2430A OM B-3

5-9SERIAL REMOTEOPERATIONThe ML2430A Series Power Meter can be operated remotely through the rearpanel serial connector (See Chapter 3, Connectors, fo

Página 238 - B-4 ML2430A OM

Requested data is returned in the same format as with GPIB, but with a preced-ing 'R' and a terminating new line character. SRQs are availab

Página 239 - ML2430A OM B-5

Chapter 5 - Procedures5-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15-2 Power Mea

Página 240 - B-6 ML2430A OM

Connected Device Character SequenceComputer connected directly "+++at\r\rS\n"Modem offline from phone networkSequence will be seen if remote

Página 241 - ML2430A OM B-7

Modem Compatibility andCommandsThe ML2430A Series Power Meter firmware supports Hayes-compatible mo-dems. The commands used are as follows:Command Def

Página 242 - B-8 ML2430A OM

4. To set a limit for channel 1 and to have the power meter autodial a remotecomputer when this limit fails, send the following sequence:LLIM 1,-12DBM

Página 243 - ML2430A OM B-9

a. Read the status register using the equivalent of a GPIB serial poll. Send themessage:!SPLNote: There is NO terminator to this message.The power met

Página 244 - B-10 ML2430A OM

To view the time profile of a signal, enter the PROFILE mode viaSYSTEM|SETUP|MODE (toggles through READOUT, PROFILE and POWER vs.TIME ). Parameters ne

Página 245 - ML2430A OM B-11

NOTEWith a CONTINUOUS trigger such as this, there will mostlikely not be specific points of interest, so the movementof the cursors is rather arbitrar

Página 246 - B-12 ML2430A OM

cussed more fully in the Triggered Measurements sectionbelow.Range Hold may be selected (see page 4-6) to limit dy-namic range and prevent small range

Página 247 - ML2430A OM B-13

The unique method of range changing applied in thismode means that the change between range 1 and 2 iseffected in less than 2 ms. In most cases it is

Página 248 - B-14 ML2430A OM

NOTEFor smaller values of display trigger delay, it is possiblethat the display will cover time intervals (on the left of thedisplay) for which there

Página 249 - ML2430A OM B-15

(a) CONNECT points. With this ON (default) the data points are connected withvectors to resemble a real time trace. When OFF, the data points are disp

Página 250 - HP-IB SUPPORT APPENDIX B

Appendix A - SpecificationsA-1 Introduction A-1A-2 System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 251 - APPENDIX B HP-IB SUPPORT

Frequency SweepModeWhen the sensor/cal factor source is set to V/GHz in Source Sweep mode, thestart and stop voltages are assumed to be 0 and 10V, and

Página 252

System|Source Sweep|Data Hold can be used to select the way in which data isplotted. Using Min/Max variation (both minimum and maximum) can be shown o

Página 253

The data can be displayed as a maximum value only, a minimum value only, max-imum and minimum values, the average of all the readings during the time

Página 254 - Menu Maps

Cal factor tables are accessed through the Sensor|CalFactor|USE TABLE frontpanel menus (Chapter 4), or through GPIB commands (Chapter 6).Example Proce

Página 255 - SENSOR MENU APPENDIX C

5-15OPTIMIZINGREADINGSThis section presents information on how to get the fastest readings from theML2430A Series power meter when operating under GPI

Página 256 - APPENDIX C SENSOR MENU

Send(0, 13, “O 1”, 3L, NLend);Receive(0, 13, buffer[i], 20, STOPend);}Settling (%) Power Level (dBm) Readings/Second0.1 0 1600.1 –30 16010.0 –30 160Th

Página 257

UsingBufferedRequestsUsing the buffered Output channel ON command (page 6-64), even faster meas-urement speeds can be achieved. By using the ON comman

Página 258

AN EXAMPLEPROGRAM IN C#include <stdio.h>/* include the NI 488.2 GPIB include file */#include “DECL.H”/* LINK with MCIB.LIB *//* Compiled with Bo

Página 259 - CHANNEL MENU APPENDIX C

Chapter 6GPIB Operation6-1INTRODUCTIONThis chapter provides alphabetically-ordered listings and descriptions of allML2430A Series GPIB programming com

Página 260 - APPENDIX C CHANNEL MENU

HP and ML4803 emulation commands on the other hand, do not have to have aspace between the command header and the parameter, or commas betweenthe para

Página 261 - TRIGGER MENU APPENDIX C

Chapter 1General Information1-1SCOPE OF THISMANUALThis manual provides installation and operation information for the ModelML2430A Series of ANRITSU P

Página 262 - APPENDIX C SYSTEM MENU

All of the Status enable type commands (*SRE for example)Stored numbers (i.e., 0, 1, 2, 3, 4, 5)Offset table numbers (i.e., 1, 2, 3, 4, 5, 6...)GPIB a

Página 263 - SYSTEM MENU APPENDIX C

“1/15/98"”Save “”cal_file"" now."'Save “cal_file” now.'<Arbitrary ASCII> This notation represents undelimited 7-bi

Página 264

6-5GPIB PC CARDSETUPThe following GPIB driver configuration set up is recommended for reliable GPIBcommunication with the ML2430A Series power meter.

Página 265

6-6USING 488.1 GPIBIEEE 488.1 level commands are in the form of data byte codes with the attention(ATN) line set. A separate function is normally prov

Página 266

Typical device library calls are 488.1 'ibrsp' and 488.2 'ReadStatusByte'.6-7USING 488.2 GPIBThe IEEE 488.1 GPIB standard was upda

Página 267

OPCOperation Complete. When a program message that includes the *OPC com-mand has been completed, and the GPIB interface is idle, with any responsesre

Página 268 - APPENDIX C CAL/ZERO MENU

2. Request data from display channel 1 by sending:O1The SRQ will be set with the new reading which will now be in the output bufferready to be read. T

Página 269

ESB If any of the event register bits are set and the corresponding event status enablebits are set the ESB bit in the status byte will be set. This b

Página 270

SENSOR The SENSOR group commands select the data acquisition controls for the se-lected sensor.SYSTEM The SYSTEM group commands control the overall fu

Página 271

6-10ML24XXA NATIVECOMMANDSThis section provides an alphabetical listing of the GPIB commands (mnemonics)used to program the Model ML2430A Series Power