1. INTRODUCTION
This manual provides maintenance instructions for
the Site Master Model S114C and S332C Antenna,
Cable, and Spectrum Analyzers. It describes the
product and provides performance verification proce
-
dures, parts replacement procedures, and a replace
-
able parts list.
2. DESCRIPTION
The Site Master is a hand held SWR/RL (standing
wave ratio/return loss), Distance-To-Fault and spec
-
trum analysis measurement instrument. It combines
a synthesized source, VSWR Bridge, receiver and
spectrum analyzer circuitry in a compact
instrument. An optional power monitor is also avail
-
able.
The following sections contain tests that can be used
to verify the performance of the Site Master models
S114C and S332C having any version of firmware.
3. SITE MASTER FREQUENCY ACCURACY
The following test can be used to verify the CW fre-
quency accuracy of the Site Master. Measurement
calibration of the Site Master is not required for this
test.
a. Equipment Required:
· Spectrum Analyzer Anritsu Model
MS2663C or equivalent
b. Procedure:
1. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
NOTE
Before continuing, allow a five minute warm
up for the internal circuitry to stabilize.
2. Press the
FREQ/DIST key, then press the
F1 soft key and set F1 to 1000 MHz, then
press the
ENTER key.
3. Press the
F2 soft key, set F2 to 1000 MHz,
then press the
ENTER key.
4. Connect the RF cable from the Site Mas
-
ter Reflection Test Port to the RF Input
on the MS2663C or equivalent.
5. Set up the Spectrum Analyzer as follows:
(a) Press the
Preset key, then select Pre
-
set All
(F1).
(b) Press the
Frequency key.
(c) Press the
1 key and then the GHz key
to change the Center Frequency to 1
GHz.
(d) Press the
Span key.
(e) Press the
7, 5, 0, and kHz keys se
-
quentially to change the Frequency
Span to 750 kHz.
(f) Press the
RBW key.
(g) Press the
1, 0 and kHz keys sequen
-
tially to change the RBW to 10 kHz.
(h) Press the
VBW key.
(i) Press the
Filter Off soft key (F3) to
turn the VB filter off.
(j) Press the
Amplitude key.
(k) Press the
0, and dBm keys sequen
-
tially to change the Reference Level
to 0 dBm.
(l) Press the
Log Scale soft key (F5)
(m) Select
2 dB/Div (F3) and the press the
return soft key (F6).
(n) Press the
Marker key.
(o) Press the
Zone Width soft key (F5).
(p) Select the
Spot soft key (F1).
6. On the Site Master, press the
SYS key,
the
OPTIONS soft key and then the FIXED
CW
soft key to turn Fixed CW on.
NOTE
If the Site Master has gone into the hold mode,
press the
RUN/HOLD key to return to normal
mode.
7. On the Spectrum Analyzer, press
A, B
then the Storage soft key (F5).
8. Press the
MaxHold soft key (F2).
9. When a smooth peak response appears
on the Spectrum Analyzer, press the
Marker Peak Search key on the Spectrum
Analyzer. Verify that the marker peak
readout value is 1000 MHz ±75 kHz.
10. On the Site Master, press the
SYS key,
the
OPTIONS soft key and then the FIXED
CW
soft key to turn Fixed CW Off.
Site Master S114C/S332C MM 1