In-band/Out-of-Channel Measurements
The in-band/out-of-channel measurements are those measurements that measure distortion
and interference within the system band, but outside of the transmitting channel. These
measurements include (1) in-band spurious emissions and (2) adjacent channel power ratio
(also called spectral regrowth). There are stringent regulatory controls on the amount of in
-
terference that a transmitter can spill to neighboring channels. In order to determine compli
-
ance with the allowable level of spurious emissions, two parameters need to be specified:
r
Measurement channel bandwidth
r
Allowable level of spurious emissions
In-band Spurious Measurement
Required Equipment
r
Site Master S114C/S332C
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator,
Anritsu 42N50A-30
r
Test Port extension cable, Anritsu 15NNF50 – 1.5C
Procedure
Step 1. Reset the Site Master to the factory default settings if a known starting state is
desired.
If the unit is off, hold down the ESC/CLEAR key while pressing the ON/OFF
key.
If the unit is already on, press the RECALL SETUP key on the data keypad.
When the key is pressed, a Recall Trace selection box appears on the display.
Select setup location 0 to recall the factory preset state.
Step 2. Using the test port extension cable and 30 dB, 50 watt, Bi-directional attenuator,
connect the Site Master to appropriate transmit test port.
Step 3.
Press the FREQ/DIST key and select the
CENTER soft key. Enter the center
frequency.
Step 4.
Press the FREQ/DIST key and select the
SPAN soft key. Set the span wide
enough to include the primary channel bandwidth and upper and lower channel
bandwidths.
Step 5.
Press the AMPLITUDE key and select the
REF LEVEL soft key to set the refer
-
ence level to 0 dBm.
Step 6. Press the
ATTEN soft key and set the attenuation to AUTO.
Step 7.
Press the SWEEP key and select the
RBW and VBW soft keys to set the resolu
-
tion bandwidth to 10 kHz and the video bandwidth to 300 Hz.
Step 8.
Press the MARKER key and press the
M1 soft key.
Step 9. Press the
EDIT soft key. Use the Up/down arrow key to move the marker over
one of the spurs. Press the ENTER key to set the marker.
Step 10. Compare the value of the marker to the specified allowable level of
in-band/out-of-channel spurious emissions for the corresponding channel trans
-
mit frequency.
5-25
Chapter 5 Spectrum Analyzer Measurements