(g) Press the 1, 0 and kHz keys sequen
-
tially to change the RBW to 10 kHz.
(h) Press the
VBW key.
(i) Press the
Filter Off soft key (F3) to
turn the VB filter off.
(j) Press the
Amplitude key.
(k) Press the
1, 0, and dBm keys sequen
-
tially to change the Reference Level
to 10 dBm.
(l) Press the
Log Scale soft key (F5)
(m) Select
2 dB/Div (F3) and the press the
return soft key (F6).
(n) Press the
Marker key.
(o) Press the
Zone Width soft key (F5).
(p) Select the
Spot soft key (F1).
6. On the Site Master, press the
SYS key,
the
OPTIONS soft key and then the FIXED
CW
soft key to turn Fixed CW on.
NOTE:
If the Site Master has gone into
the hold mode, press the
RUN/HOLD key to return to normal
mode.
7. On the Spectrum Analyzer, press the
A, B
and then the Storage key (F5).
8. Press the
Max Hold soft key (F2).
9. A peak response should appear on the
Spectrum Analyzer.
10. Press the
Marker Peak Search key on the
Spectrum Analyzer. Verify that the
marker peak readout value is 1000 MHz
±75 kHz.
11. On the Site Master, press the
SYS key,
the
OPTIONS soft key and then the FIXED
CW
soft key to turn Fixed CW Off.
5. RETURN LOSS VERIFICATION
The following test can be used to verify the accuracy
of return loss measurements. Measurement calibra
-
tion of the Site Master is required for this test.
a. Equipment Required:
·
20 dB offset, Anritsu SC5270
·
6 dB offset, Anritsu SC5237
·
Open/Short, Anritsu 22N50
·
50 Ohm Termination,
Anritsu 28N50-2 or SM/PL
b. Procedure:
1. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
NOTE
Before continuing, allow a five
minute warm up for the internal
circuitry to stabilize.
2. Press the
MODE soft key.
3. Use the Up/Down Arrow key to highlight
RETURN LOSS, then press ENTER.
4. Press the
START CAL key.
5. Follow the instructions on the screen to
perform a calibration using a 22N50
Open/Short and 28N50-2 or SM/PL Ter
-
mination.
6. Connect the 20 dB offset to the Refl Test
Port and verify that the reading is:
q
S113C/S115BQ: 20 dB ± 1.7 dB
q
S331C: 20 dB ± 1.7 dB
7. Connect the 6 dB offset to the Refl Test
Port and verify that the reading is:
q
S113C/S115BQ: 6 dB ± 1.2 dB
q
S331C: 6 dB ± 1.2 dB.
Site Master S113C/S331C/S115BQ MM 3