SECTION 6 GPIB
6-30
:FREQuency:SWEep:SPAN <Numeric> <Freq term>
Function Sets the span frequency for a sweep.
Parameter <Numeric> = 0.02 Hz to 1040 MHz (MG3641A)
0.02 Hz to 2080 MHz (MG3642A)
Unit <Freq term>
Restriction None
:FREQuency:SWEep:SPAN?
Function Reads out the span frequency for a sweep.
Response Span frequency for a sweep
Restriction None
:FREQuency:SWEep:STEP:SIZE <Numeric> <Freq term>
Function Sets the frequency step size for a liner sweep.
Parameter <Numeric> = 0.01 Hz to 1040 MHz (MG3641A)
0.01 Hz to 2080 MHz (MG3642A)
Unit <Freq term>
Restriction None
:FREQuency:SWEep:STEP:SIZE?
Function Reads out the frequency step size for a liner sweep.
Response Frequency step size for a liner sweep
Restriction None
:FREQuency:SWEep:STEP:NUMBer <Numeric>
Function Sets the number of frequency points for a liner sweep.
Parameter <Numeric> = 2 to 104000000001 (MG3641A)
2 to 208000000001 (MG3642A)
Unit <Non term>
Restriction None
:FREQuency:SWEep:STEP:NUMBer?
Function Reads out the number of frequency points for a liner sweep.
Response Number of frequency points for a liner sweep
Restriction None
:FREQuency:SWEep:MODE <Character>
Function Selects the frequency sweep mode.
Parameter <Character> = AUTO
SINGLE
MANUAL
Unit <Non term>
Restriction None
:FREQuency:SWEep:MODE?
Function Reads out the frequency sweep mode.
Response Frequency sweep mode = AUTO
SINGLE
MANUAL
Restriction None