Anritsu ME7840A Manual do Utilizador Página 48

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 126
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 47
NOTE
The linear power calibration takes up-to-5 minutes to
complete. It calibrates the output power at the refer
-
ence plane for both internal sources (power out of
Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci
-
fication file. When the test completes, the “Next” but
-
ton will become available.
Step 5. Press “Next” to calibrate flat power response (below).
NOTE
The flat power calibration also takes less than 1 min
-
ute to complete. It calibrates the output power at the
reference plane for both internal sources (power out
of Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci
-
fication file. When the test completes, the “Next” but
-
ton will become available.
POWER LEVEL OPERATIONS, CALIBRATION
4-6 ME 7840A OMM
Vista de página 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 125 126

Comentários a estes Manuais

Sem comentários