bit 7 : Not Used
168) Status Byte 7: (0b = Off, 1b = On)
(LSB) bits 0-1: Not Used
bit 2: DIST-RL Multiple Limit Segment 1 Status On/Off
bit 3: DIST-RL Multiple Limit Segment 2 Status On/Off
bit 4: DIST-RL Multiple Limit Segment 3 Status On/Off
bit 5: DIST-RL Multiple Limit Segment 4 Status On/Off
bit 6: DIST-RL Multiple Limit Segment 5 Status On/Off
bit 7: Not Used
169) Status Byte 8:
(LSB) bits0-1:DTFWindowing Mode
bit: 1 0
||
0 0 - Rectangular (No Windowing)
0 1 - Nominal Side Lobe
1 0 - Low Side Lobe
1 1 - Minimum Side Lobe
bit 2 : Serial Port Echo Status On/Off
bits3–7:NotUsed
170) Status Byte 9: (0b = Off, 1b = On )
(LSB) bit 0 : Fixed CW Mode On/Off
bit 1 : Site Master VNA Cal On/Off
bit 2 : LCD Back Light On/Off
bit 3 : Measurement Unit Metric/English (0b = English, 1b = Metric)
bit 4 : InstaCal On/Off
bits 5-6: Not Used
bit 7 : Cal Mode (0b = OSL Cal, 1b = FlexCal)
171) VNA Signal Standard
96
(higher byte)
172) VNA Signal Standard (lower byte)
173-300) Not Used
For Spectrum Analyzer and Transmission Measurement (Option 21) Modes:
26) Spectrum Analyzer Mode Data Points (higher byte)
27) Spectrum Analyzer Mode Data Points (lower byte)
28) Spectrum Analyzer Start Frequency
97
(highest byte)
29) Spectrum Analyzer Start Frequency
30) Spectrum Analyzer Start Frequency
31) Spectrum Analyzer Start Frequency (lowest byte)
32) Spectrum Analyzer Stop Frequency (highest byte)
33) Spectrum Analyzer Stop Frequency
34) Spectrum Analyzer Stop Frequency
35) Spectrum Analyzer Stop Frequency (lowest byte)
36) Spectrum Analyzer Center Frequency (highest byte)
37) Spectrum Analyzer Center Frequency
38) Spectrum Analyzer Center Frequency
39) Spectrum Analyzer Center Frequency (lowest byte)
40) Spectrum Analyzer Frequency Span (highest byte)
41) Spectrum Analyzer Frequency Span
42) Spectrum Analyzer Frequency Span
43) Spectrum Analyzer Frequency Span (lowest byte)
44) Spectrum Analyzer Minimum Frequency Step Size (highest byte)
40 Site Master PM
96 Index into Standard List (use control byte #89 to retrieve the ASCII string name). “No Standard” sent as FFFEh
97 Scaled by frequency scale factor (bytes 321-322)
Vista de página 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 131 132

Comentários a estes Manuais

Sem comentários

Baldwin 3861.112.18 Handbücher

Bedienungsanleitungen und Benutzerhandbücher für Nein Baldwin 3861.112.18.
Wir stellen 1 PDF-Handbücher Baldwin 3861.112.18 zum kostenlosen herunterladen nach Dokumenttypen zur Verfügung Installationsanleitung






Weitere Produkte und Handbücher für Nein Baldwin

Modelle Dokumententyp
5015 112 PASS 2-3/8 SET Installationsanleitung   Baldwin 5015 112 PASS 2-3/8 SET Installation Guide, 1 Seiten
8232.003 Gebrauchs- und Pflegehandbuch   Baldwin 8232.003 Use and Care Manual, 1 Seiten
9BR1801-696 Bedienerhandbuch    Baldwin 9BR1801-696 Instructions / Assembly [no] , 1 Seiten
180WYHXATK RDB L03 SMT CP Gebrauchs- und Pflegehandbuch    Baldwin 180WYHXATK RDB L03 SMT CP Use and Care Manual, 2 Seiten
SC.NAPxROU.TRR.112 Bedienerhandbuch    Baldwin SC.NAPxROU.TRR.112 Instructions / Assembly, 2 Seiten
183SPEXSPL SQR 15 SMT CP Bedienerhandbuch    Baldwin 183SPEXSPL SQR 15 SMT CP Instructions / Assembly, 2 Seiten
3865.150 Installationsanleitung     Baldwin 3865.150 Installation Guide, 1 Seiten