Anritsu ME7840A Especificações Página 135

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 134
6. Press the menu soft key to select the source to be calibrated:
Select Source 1/2
7. Connect the power sensor to the selected port (Port 1 for Source 1,
Port 3 for Source 2).
8. Select:
START CAL
9. After the ALC calibrations are complete, back up the calibrations by
selecting SAVE TO HARD DISK.
If the calibration fails:
q
Verify that the power meter and sensor are functioning correctly,
the GPIB cables are in good condition, and all connections are se
-
cure.
q
Note the failed step in the calibration procedure from the screen
message.
q
Repeat the calibration.
If it still fails, proceed to Chapter 6—Troubleshooting.
ADJUSTMENTS ALC ADJUSTMENT PROCEDURE
MS462XX MM 5-9
Vista de página 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 204 205

Comentários a estes Manuais

Sem comentários